How do I choose audit filters for my opacity monitor?

Ongoing quality assurance is important for any piece of measuring equipment. From a voltmeter to a spectrophotometer, calibration checks are needed to demonstrate that it is continuing to function within its design specification. An opacity monitor is no different, and the usual method for verification is to perform a calibration error check using traceable audit filters. These are neutral density filters which are designed to have a stable opacity value, and which are periodically calibrated to a traceable national standard. A traceable calibration is important because it specifies both the filter value and its associated uncertainty.

In most cases, the required audit filter values are determined by the typical measurement range of the opacity monitor and will often be specified by a national or international standard. An opacity monitor used for compliance measurements in the USA must comply with the ASTM D6216 standard. This specifies low-, mid- and high-level values depending on the emission limit value for the installation.

Emission Limit Value    <20% Opacity     >20% Opacity 
 Low  5% to 10%  10% to 20%
 Mid  10% - 20%  20% to 30%
 High  20% to 405%   30% to 40%

Note that the specified values are the instrument reading after any applicable pathlength correction factor (PLCF) has been applied. Filter suppliers and calibration labs can calculate the required corrections, but it is a good idea for the instrument user to verify the correct value. The following equations allow you to calculate the relevant value. If Opacitycorr is the value in the table above and the required filter value is Opacityreq

ODcorr = -log10 (1 – Opacitycorr)

Opacityreq = 1 – 10(-PLCF x ODcorr)
 
Where opacity values are expressed as decimal values. For example, an installation with an ELV of 10% and PLCF = 0.800 requires a low-level filter value between 5% (decimal 0.05) and 10% (decimal 0.10). The minimum filter value is calculated as:

ODcorr = -log10 (1 – 0.05) = 0.022

Opacityreq = 1 – 10(-0.022 x 0.800) = 1 – 0.960 = 0.040 = 4.0%

 The maximum filter value is calculated as:

ODcorr = -log10 (1 – 0.10) = 0.046

Opacityreq = 1 – 10(-0.046 x 0.800) = 1 – 0.919 = 0.081 = 8.1%

Therefore the required value on the calibration certificate must be between 4.0% and 8.1%.

In conclusion, periodic audits are essential to verify the correct operation of your opacity monitor, and you should now understand how to choose the correct audit filter values for your opacity monitor. If you have any questions, please  contact us. 

Archives

Skip Navigation Links.
Collapse 2023(25)2023(25)
Collapse November(4)November(4)
Importance of Reliable Slag Detection
How to Enhance Combustion with NIR-Bs to Reduce Emissions and Boost Energy Efficiency
Ensuring Quality and Control in Auto-pour Processes for Foundry Castings
Can I Measure Opacity When I Have Water in my Flue Gas?
Collapse October(2)October(2)
What is EtherNet/IP and Why Do I Need It?
How Critical is Measurement of Temperatures in the Aluminium Industry?
Collapse August(1)August(1)
Industrial Thermal Imaging Benefits in Heat Treatment Applications
Collapse July(1)July(1)
Why Should I Use Two NIR-Bs on Container Glass Furnaces?
Collapse June(1)June(1)
Exploring Career Opportunities at AMETEK Land: Your Pathway to Success
Collapse May(3)May(3)
SPOT AL: Revolutionising Aluminum Extrusion with Precision Temperature Control
5 Practical Tips for Using the Gold Cup to Measure Tube Wall Temperatures
Making Smart Investments in Non-Contact Temperature Measurement Technology
Collapse April(3)April(3)
INDUSTRY 4.0 & 5.0 in Non-contact Temperature Measurement Applications and Devices
Gas Concentration Measurements in Glass Melt Tanks with a Portable Gas Analyser
How to Use the New LWIR-640 and its IMAGEPro Software to Monitor the Exterior of a Glass Furnace and Avoid Glass Leaks
Collapse March(3)March(3)
What is Dual-Sensor Technology?
Maintaining the Reliability and Safety of Steam Methane Reformers through Outlet System Monitoring
Combining a Transportable NIR-B-2K & Lancom 4 for Thermal Surveys & Combustion Optimisation
Collapse February(4)February(4)
Using Fiber-Optic Instruments (FGs) for Live Temperature Monitoring in Container Glass Furnaces
Combustion Control in a Black Liquor Boiler
Benefits of Using Medium-wavelength (MWIR) Thermal Imagers Instead of Widely Used Long-wavelength (LWIR) Cameras
10 Reasons Why Thermal Cameras Are Preferred to Video Cameras for Furnace Monitoring.
Collapse January(3)January(3)
Steam Methane Reformer (SMR) Tube Metal Temperatures (TMT) Data Collection and Correction
How to Improve Temperature Gradient and Quality of Glass Ribbons Using The LSP-HD Linescanner
Understanding the Importance of 'HOT SPOT' Measurement in Non-Contact Temperature Measurement Applications
Expand 2022(30)2022(30)
Expand 2021(19)2021(19)
Expand 2020(15)2020(15)
Expand 2019(15)2019(15)
Expand 2018(9)2018(9)